A medical device’s maintenance history provides a record of the maintenance, repair and inspection procedures carried out during its use. Up-to-date documentation makes it easier to monitor the condition and reliability of the device and to assess future maintenance needs.
Maintenance history becomes particularly important when a device develops recurring faults or when its remaining service life and repair needs are being assessed. Documentation makes it possible to review previous procedures, monitor the development of faults and demonstrate that the device has been properly maintained.
In this article, we explain what should be included in a medical device’s maintenance history, what Finnish legislation and Fimea’s guidance require, and how maintenance reporting supports lifecycle management.
Why is medical device maintenance documentation important?
Maintenance documentation provides the foundation for tracking a medical device’s maintenance history. It makes it possible to review previously identified faults, repairs and maintenance procedures, as well as how the condition of the device has developed over time.
An up-to-date maintenance history also helps with planning future maintenance and identifying changes in device reliability. Recurring faults, shorter maintenance intervals or repeated repairs to the same components may indicate that the device requires closer attention or that replacement should be considered.
Documentation also improves traceability. Previous procedures can be reviewed when investigating a new fault, deviation or adverse incident. At the same time, the records provide evidence of what was done to the device, when it was done and why.
What is required of medical device maintenance documentation?
There is no single documentation format that every organisation must follow. However, the documentation must support device traceability and make it possible to demonstrate that the device has been properly maintained and serviced. In Finland, key requirements are based on medical device legislation and Fimea’s guidance for professional users.
Fimea’s guidance on documenting medical device maintenance
The Finnish Medical Devices Act requires professional users to ensure that devices are adjusted, maintained and serviced in accordance with the manufacturer’s instructions and otherwise appropriately. Organisations must also be able to demonstrate that these obligations have been fulfilled. According to Fimea, organisations should therefore have a procedure in place for verifying that maintenance and servicing have been carried out in accordance with the manufacturer’s instructions.
Fimea has specifically noted that legislation does not prescribe one particular method for recording maintenance activities. Documenting maintenance history is nevertheless an important part of quality management, as it provides evidence of the maintenance and repair procedures performed on a device and of how appropriate maintenance has been ensured.
According to Fimea, maintenance history may also be held by an external maintenance service provider. The professional user must nevertheless have access to the information when necessary, and the maintenance history must be available, for example, when investigating an adverse incident.
ISO 13485 and maintenance documentation
ISO 13485 is an international standard for quality management systems for medical devices. From a maintenance perspective, the standard requires servicing activities to be carried out according to documented procedures and appropriate records to be retained for the work performed.
The methods, reference materials and measurements used during maintenance must make it possible to verify that the device meets the applicable requirements after servicing. In practice, sufficient documentation should therefore remain to show what was done to the device and how its functionality was verified.
Under ISO 13485, maintenance records are used to monitor the maintenance history of the device and to identify recurring faults and other potential quality issues. The information can be used to determine whether further investigation or corrective action is required.
Maintenance records must also be controlled so that they can be located when needed and linked to the correct device. The purpose of the documentation is to support traceability and demonstrate that maintenance procedures have been carried out in accordance with the defined process.
What should be recorded in a medical device’s maintenance history?
The maintenance history should contain sufficient information to verify the work performed and link it to the correct device. The documentation should clearly identify the device, explain why maintenance was required, describe the procedures carried out and show how the device’s functionality was verified afterwards.
Device identification details
Maintenance records must be clearly linked to the correct device. The documentation should therefore include the information required to identify it, such as the manufacturer, model, serial number or another device identifier used by the organisation.
Date and reason for maintenance
The documentation should show when the maintenance was performed and what prompted it. This may include scheduled preventive maintenance, an identified fault, an inspection or another maintenance requirement.
Identified faults and deviations
Faults and other deviations detected during maintenance should be recorded as part of the maintenance history. This makes it possible to review the condition of the device before servicing and determine whether the same problems have occurred previously.
Maintenance and repair procedures performed
The maintenance history should provide evidence of the procedures carried out on the device. Records may include repairs, adjustments, inspections and other maintenance activities.
Verification of device functionality
Post-maintenance documentation should show how the functionality of the device was checked before it was returned to service. The required testing and inspections depend on the type of device and the maintenance procedures performed.
Person or organisation performing the maintenance
The person or organisation responsible for the maintenance should also be recorded. This makes it possible to trace who carried out the work and to request further information about the maintenance later if necessary.
Where should medical device maintenance history be stored?
Maintenance history may be stored in an organisation’s own equipment management system, in maintenance reports or in the system of an external maintenance service provider. The essential requirement is that the information remains accessible when needed and can be clearly linked to the correct device.
Access to maintenance information is particularly important when investigating a new fault, deviation or adverse incident. Previous records should make it possible to determine what procedures were carried out on the device and when they were performed.
Using an external maintenance service provider does not remove the professional user’s responsibility for ensuring that the device is appropriately maintained. Organisations should therefore ensure that maintenance histories and reports are available to them when required.
How does Health Traders document completed maintenance?
Health Traders always provides the customer with a maintenance report after servicing or repairing a device. The report includes the necessary and legally required information relating to the maintenance of the device, as well as details of the repairs, inspections, adjustments and maintenance procedures performed.
The maintenance report forms part of the device’s maintenance history and helps the customer monitor the procedures carried out throughout the device’s lifecycle. Previous maintenance activities can later be reviewed and the information used, for example, when assessing a new maintenance requirement or investigating a recurring fault.